Analysis and Comparison of Data of Linear Accelerators Measured with Diode and Ionization Detector

孙文钊,张丹丹,邓小武
DOI: https://doi.org/10.3969/j.issn.1005-202X.2014.01.008
2014-01-01
Abstract:Objective: This study was to analyze and compare the data measured with ionization chambers and diode detector in three dimensional water tanks and provided theory basis for foundation of precise calculating model in treatment planning system. Methods: The treatment planning system needs plenty of accurate physical data to model treatment machine. These data include the total scatter factors, percent depth dose curves, off-axis ratio curves and so on. These data were acquired in linear accelerator of 8 MV photon beam with different beam fields. Results: For total scatted factors,the great discrepancy exist in small fields, for field of 1cm×1cm, the discrepancy from two detectors is 15.32%; For percent depth dose, the great discrepancy exists in build-up region. The discrepancy for two types detectors are more than 10% for all field sizes. For off-axis ratio curves, the data showed differ in penumbra areas. The field sizes calculated by data measured from diode detector were smaller than that from ion chamber. Conclusions: For total scatter factors of small fields, diode and ion chamber with small volume can get good result. For percent depth dose curves measurement, ion chamber is recommended. For off-axis ratio curves measurement, diode detectors can performance well than ion chamber.