Lactate Biosensor Using Covalently Bonded Lactate Oxidase in Egg Membrane

邓健,廖力夫,袁亚莉,肖丹
DOI: https://doi.org/10.3969/j.issn.1000-0720.2002.06.018
2002-01-01
Abstract:Lactate oxidase was immobilized in egg membranes and crosslinked with bovine serum albumin with glutaraldehyde. The enzyme electrode was than constructed by putting the enzyme membrane onto an oxygen electrode. In phosphate buffers at pH 6.8 and at the temperature of 25~37 ℃, the response of the biosensor had a linear relationship with lactate concentration in the range of 5.0×10 -5 ~2 5×10 -2 mol/L with detection limit of 2.0×10 -5 mol/L and RSD of 1.4%( n =11). The biosensor was used for clinical assay of lactate in plasma and other body fluids.