Comparison of Performance of Lanthanum Bromide, Sodium Iodide and Plastic Scintillator Detectors

Zheng-wei GU,Wen-qi YAN,Wen-bin LIU,Gao-long ZHANG
DOI: https://doi.org/10.3969/j.issn.1006-7167.2015.04.017
2015-01-01
Abstract:To wholly recognize the characteristics of scintillator detectors, the signal characteristics, the energy resolution,the time resolution and the other properties of several common scintillator detectors( lanthanum bromide,sodium iodide and plastic scintillation detectors) are measured by using60 Co and137Cs gamma sources as well as90Sr-90 Y beta source in this experiment,and the results are compared. It shows that the plastic scintillator detector is suitable for measuring time signal. Sodium iodide detector is suitable for energy measurement,lanthanum bromide detector which lies between the two detectors mentioned above,can be used for both time measurement and energy measurement. It is a new kind of scintillator detector and will be gradually used in the future scientific research.