Performance comparison of four compact room-temperature detectors-two cadmium zinc telluride (CZT) semiconductor detectors, a LaCl/sub 3/(Ce) scintillator, and an NaI(Tl) scintillator

J.K. Hartwell,R.J. Gehrke,M.E.M. Ilwain
DOI: https://doi.org/10.1109/tns.2005.856599
IF: 1.703
2005-10-01
IEEE Transactions on Nuclear Science
Abstract:The performance characteristics of four compact, room-temperature detectors—two scintillators and two semiconductor detectors—were studied. All are commercially available detectors. The two scintillators were a ø13 mm×13 mm lanthanum chloride [LaCl${}_{3}$(Ce)] detector and a ø25 mm×25 mm sodium iodide [NaI(Tl)] detector. The two semiconductor detectors were a 10×10×3 mm${}^{3}$ cadmium zinc telluride (CZT) detector with a coplanar gridded anode and a 5×5×5 mm${}^{3}$ CZT detector with an extended cathode. The efficiency, resolution, and peak shape of these devices are compared. Since LaCl${}_{3}$(Ce) is a relatively new commercial scintillator material, additional information on the performance of this detector is presented. Specifically, the impact of naturally occurring radioactive ${}^{138}$ La and additional contamination from alpha-emitting radionuclides on the background measured with this detector are discussed.