Study of Liver Volume Measurement and Its Clinical Application for Liver Transplantation Using Multiple-Slice Spiral CT

彭志毅,俞哲锋,黄东胜,邝平定,郑树森,肖圣祥,吴健
DOI: https://doi.org/10.3760/j.issn:1005-1201.2004.03.015
2004-01-01
Abstract:Objective To study the accuracy of liver volume measurement using MSCT and its application in liver transplantation.Methods (1) Experimental study.Ten pig livers were scanned using MSCT with two collimations (3.2 mm and 6.5 mm) and pitch 1.25.Semi-automatic method was used to reconstruct 3D liver models to measure the liver volume.(2) Clinical study.Twenty-three patients received MSCT scan with collimation of 6.5 mm before liver transplantation.Same method was used to calculate the liver volume and the measurement was repeated by the same observer after 1 month.Results (1) Experimental study.Actual liver volumes were (1134.1±288.0) ml.Liver volumes by MSCT with two collimations were (1125.0±282.5) ml (3.2 mm) and (1101.6±277.6)ml (6.5 mm).The accuracy was (99.5±0.8)% and (97.4±0.8)%,respectively.Both showed same good agreement with actual liver volume: r=0.999, P0.01.(2) Clinical study.Actual liver volumes were (1455.7±730.0) ml.Liver volume by MSCT was (1462.7±774.1) ml.The accuracy was (99.5±9.6)%,r=0.986,P0.01.Liver volume measured again was (1449.4±768.9) ml,r=0.991 (P0.01).Conclusion MSCT can assess the liver volume correctly,and could be used as a routine step for evaluations before liver transplantation.
What problem does this paper attempt to address?