Determination of Creatinine and Uric Acid in Urine with Direct Sample Injection by High Performance Liquid Chromatography

Song Yanxi,Zhang Haixia,Liu Mancang
DOI: https://doi.org/10.3321/j.issn:0253-3820.1999.02.029
1999-01-01
Abstract:A direct sample injection approach for the determination of creatinine and uric acid in urine has been developed. The chromatographic separation is carried out on a 3-aminopropyl crosslinking polyoxyethylene mixed functional phase columm,where proteins in biological fluids are not retained.The mobile phase consists of 0.01mol/L phosphate buffer and 10% methanol with pH 4. 81 UV detector is used at 235 nm.Uric acid and creatinine have linear response ranges of 3ng/mL-10μg/mL and 2ng/mL-10μg/mL,respectively.The recoveries of uric acid and creatinine are 96.3%(C.V.=2.0%) and 102.0%(C.V=3.0%),respectively.The total run time is not more than 5 minutes.It is useful for clinical purpose.