Abstract:In clinical medicine, urine creatinine concentration is an important marker in the evaluation of renal function and muscular dysfunctions. Herein, we reported a novel method for rapid determination of creatinine in urine by microchip electrophoresis with light-emitting diode induced fluorescence detection. Creatinine was derivatized by fluorescein isothiocyanate, and then quantitatively detected by the developed microchip LED induced fluorescence detection system. The excitation and emission wavelengths were 490 and 523 nm, respectively. The urine samples were analyzed after centrifuge and filtration. A baseline separation was obtained in <30 s using 10 mM borate buffer (pH 9.0, containing 45 mM sodium dodecylsulfate), with separation voltage of 1.5 kV. Good linearity was obtained (r 2 = 0.9978) in the concentration range of 10.0–2.00 × 103 μM, and the limit of detection was 2.87 μM (S/N = 3). The recovery was 96.0–107 %, and the interday precision was <4.5 % (n = 6). To validate assay results, we compared the present method with the Jaffe’s colorimetric assay by measuring real urine samples. The method was reliable, sensitive, high-speed, low-cost and suitable for the routine analysis of creatinine in biofluids.