Direct determination of creatinine based on poly(ethyleneimine)/phosphotungstic acid multilayer modified electrode

ping han,shimei xu,shun feng,yanjun hao,jide wang
DOI: https://doi.org/10.1016/j.talanta.2016.01.017
IF: 6.1
2016-01-01
Talanta
Abstract:In this work, the direct determination of creatinine was achieved using a poly(ethyleneimine)/phosphotungstic acid multilayer modified electrode with the assistance of Copper(II) ions by cyclic voltammetry. The quantity of creatinine were determined by measuring the redox peak current of Cu(II)-creatinine complex/Cu(I)-creatinine complex. Factors affecting the response current of creatinine at the modified electrode were optimized. A linear relationship between the response current and the concentration of creatinine ranging from 0.125 to 62.5μM was obtained with a detection limit of 0.06μM. The proposed method was applied to determine creatinine in human urine, and satisfied results were gotten which was validated in accordance with high performance liquid chromatography. The proposed electrode provided a promising alternative in routine sensing for creatinine without enzymatic assistance.
What problem does this paper attempt to address?