Postoperative Follow-up of 16 Patients with Implantable Cardioverter-Defibrillator

赵玲,郭涛,韩明华,肖春晖,刘中梅,李淑敏,赵文盛,粟建国,柳永华,李佳
DOI: https://doi.org/10.3969/j.issn.1007-2659.2003.05.008
2003-01-01
Abstract:16 patients,male 7 and femal 9,mean age 49 years with implantable cardioverter-defibrillator(ICD)were followed up in order to give them an approprate treatment by changing the programsetting .16 patients with malignant ventricular tachycardia (VT) underwent implantation of ICD with multitherapies including antitachycardia pacing (ATP),cardioversion,defibrillation and data storing function.The working program of ICD was set up according to the results of electrophysiologic study.Follow-up evaluation of patients after ICD implantation was conducted before discharge,after the first spontaneous arrhythmia event and every 3 months after discharge.Result: In the follow-up of 30 months,the storage data revealed that 789 episodes of VT were detected and 127 of them were nonsustained VT which spontaneously terminated.661(99.9%)VT episodes were treated by ICD successfully.650 (98%) were terminated by ATP and 9(1.4%) by low energy cardioversion.2 (0.3%)VT were accelerated to VF and terminated by energy shocks.There were 37 VF episodes diagnosed by the devices.All the episodes were terminated by shocks.2 patients changed ICD.The original catheter system could be used continuously. During the follow-up one patients died of heart failue and the other died of suicide.Conclusion:Careful postoperative observation and reprogramming after spontaneous event are necessary.Psychological supports and therapies should be done to the paced patients.
What problem does this paper attempt to address?