16 detector-row CT: parameters and postprocessing of portography

zhang kai,li chuanfu,ma xiangxing,li dumin,cui fengyu,wang qing,qi zihua,wang luzhong
DOI: https://doi.org/10.3969/j.issn.1671-7554.2005.02.019
2005-01-01
Abstract:Objective: To find a suitable delayed time, thickness and postprocessing way to display portal venous system. Methods: The density difference between the portal vein and liver of 55s and 65 s groups was compared. Using the raw data of 55s group, images were reconstructed with the thickness of 0.75 mm, 1.0 mm, 3.0 mm respectively. Then the imagings under MIP and MPR were compared. Results: The CT value between portal vein and liver of 55 s group was significantly higher than that of 65 s group(P<0.05). The reconstructed images of 1.0 mm was better than that of 3.0 mm, but had no significant difference with that of 0.75 mm(P>0.05). MIP reconstruction images were better than those of MPR. Conclusion: Used 55 s, and 1.0 mm thickness by MIP high quality portography can be got.