Comparison of Acute Hemodynamic Effect of Prioritizing Ventricular Resynchronization Versus Left Ventricular Filling During Optimization of Cardiac Resynchronization Therapy
Yixiu Liang,Ahran D. Arnold,Nadine Ali,Jingfeng Wang,Xue Gong,Ziqing Yu,Xi Liu,Hongyang Lu,Matthew J. Shun-Shin,Daniel Keene,Andrew M. Leong,Akriti Naraen,Weiwei Zhang,Ruogu Li,Weijian Huang,Yangang Su,Zachary I. Whinnett
DOI: https://doi.org/10.1016/j.hrthm.2024.09.012
IF: 6.779
2024-01-01
Heart Rhythm
Abstract:Background Targeting the maximal ventricular resynchronization, with the shortest QRS duration (QRSd), is commonly implemented after cardiac resynchronization therapy (CRT). Objective We compared optimization of ventricular resynchronization with optimization of left ventricular (LV) filling during CRT by measuring their acute hemodynamic effects. Methods Patients with standard CRT indications, recruited from two centers, underwent biventricular pacing (BVP) and left bundle branch pacing (LBBP). We performed a within-patient comparison of acute hemodynamic response of systolic blood pressure (SBP) at the atrioventricular delay (AVD) with the shortest QRSd against the AVD with the most efficient LV filling. In a validation sub-study, we also performed electrical assessment using QRS area (QRSa) and hemodynamic assessment with the maximum rate of LV pressure rise (dP/dtmax). Results Thirty patients (age 65 ± 10, 53% male) were recruited. The AVD producing maximal ventricular resynchronization was associated with a significantly shorter QRSd (difference 15 ± 12 ms for BVP and 18 ± 13 ms for LBBP, both P < 0.01) and a significantly smaller improvement in SBP (difference 3 ± 4 mmHg for BVP, and 2 ± 2 mmHg for LBBP, both P < 0.01) compared with the AVD that optimized filling. Similar findings were observed in the sub-study, with a significantly smaller improvement in dP/dtmax assessed with QRSd and QRSa (difference 9 ± 7% and 6 ± 4% during BVP, and 5 ± 6% and 3 ± 3% during LBBP, all P < 0.01). Conclusion Targeting the maximal ventricular resynchronization results in suboptimal acute hemodynamic performance with both BVP and LBBP as CRT. These findings support prioritizing LV filling when programming AVD for CRT.