Prospectively ECG-triggered High-Pitch 80 Kvp Coronary Computed Tomography Angiography with 30 Ml of 270 Mg I/mL Contrast Material and Iterative Reconstruction.

Li Qi,Sheng Yong Wu,Felix G Meinel,Chang Sheng Zhou,Qing,Andrew D McQuiston,Xue Man Ji,U Joseph Schoepf,Guang Ming Lu,Long Jiang Zhang
DOI: https://doi.org/10.1177/0284185115590433
2015-01-01
Abstract:BACKGROUND:The smallest diagnostically appropriate amount of contrast medium should be used in coronary computed tomography angiography (CCTA).PURPOSE:To investigate the feasibility of prospectively ECG-triggered high-pitch CCTA using 30 mL of 270 mg I/mL contrast material, 80 kVp, and iterative reconstruction (IR).MATERIAL AND METHODS:Eighty-two consecutive patients underwent CCTA with a prospectively ECG-triggered high-pitch protocol. Forty-three patients were examined at 100 kVp with filtered back projection after 60 mL of 370 mg I/mL contrast material was administered. Another 39 patients were examined at 80 kVp with IR after 30 mL of 270 mg I/mL contrast material was administered. Subjective and objective image quality was evaluated for each patient. Radiation doses were estimated and compared.RESULTS:Mean attenuation, noise and signal-to-noise ratio in 80 kVp group were significantly lower than in 100 kVp group (all P < 0.05), while there was no significant difference in contrast-to-noise ratio (CNR), although a trend towards a lower CNR in 80 kVp group was observed (P = 0.099). The subjective image quality between the two groups was not significantly different (P = 0.905). The effective dose and iodine load in 80 kVp group were reduced by 54% and 64%, respectively, when compared with 100 kVp group.CONCLUSION:Prospectively ECG-triggered high-pitch CCTA at 80 kVp with 30 mL of 270 mg I/mL contrast material and IR is feasible for patients with BMI less than 25 kg/m(2) and reduces radiation dose and iodine load when compared with the standard CCTA protocol.
What problem does this paper attempt to address?