Extended Axial Imaging Range, Widefield Swept Source Optical Coherence Tomography Angiography

Gangjun Liu,Jianlong Yang,Jie Wang,Yan Li,Pengxiao Zang,Yali Jia,David Huang
DOI: https://doi.org/10.1002/jbio.201600325
IF: 3.3899
2017-01-01
Journal of Biophotonics
Abstract:We developed a high-speed, swept source OCT system for widefield OCT angiography (OCTA) imaging. The system has an extended axial imaging range of 6.6 mm. An electrical lens is used for fast, automatic focusing. The recently developed split-spectrum amplitude and phase-gradient angiography allow high-resolution OCTA imaging with only two B-scan repetitions. An improved post-processing algorithm effectively removed trigger jitter artifacts and reduced noise in the flow signal. We demonstrated high contrast 3 mm x 3 mm OCTA image with 400 x 400 pixels acquired in 3 seconds and high-definition 8 mm x 6 mm and 12 mm x 6 mm OCTA images with 850 x 400 pixels obtained in 4 seconds. A widefield 8 mm x 11 mm OCTA image is produced by montaging two 8 mm x 6 mm scans. An ultra-widefield (with a maximum of 22 mm along both vertical and horizontal directions) capillary-resolution OCTA image is obtained by montaging six 12 mm x 6 mm scans.