Optical Coherence Tomography Angiography with Adaptive Multi-Time Interval.

Bin He,Yuxin Zhang,Zhe Meng,Zhijun He,Chengming Wang,Zhenyu Chen,Zichen Yin,Zhangwei Hu,Yejiong Shi,Ning Zhang,Wenxin Zhang,Guihuai Wang,Ping Xue
DOI: https://doi.org/10.1002/jbio.202200340
IF: 3.3899
2023-01-01
Journal of Biophotonics
Abstract:Optical coherence tomography angiography (OCTA) can provide in vivo three-dimensional microvasculature information of bio-tissues, but it is sensitive to motion and time-consuming. To overcome these limitations, we propose an adaptive multiple time interval correlation mapping OCTA with a time-efficient scanning protocol and motion compensation algorithms. A spectral-domain OCT with a center wavelength of 850 nm, A-scan rate of 120 kHz and spatial resolution of 4.1 mu m (axial) x 6.9 mu m (lateral) is built to reconstruct the microvascular networks in the human arm. By adaptive optimization of the weights of different time interval B-scan angiograms, our novel OCTA technique achieves better performance with a visible vascular density increase of similar to 67% and a signal-to-noise ratio enhancement of similar to 11.6%.
What problem does this paper attempt to address?