High Performance OCTA Enabled by Combining Features of Shape, Intensity, and Complex Decorrelation

Huakun Li,Kaiyuan Liu,Tongtong Cao,Lin Yao,Ziyi Zhang,Xiaofeng Deng,Chixin Du,Peng Li
DOI: https://doi.org/10.1364/ol.405751
2021-01-01
Abstract:Motion contrast optical coherence tomography angiography (OCTA) entails a precise identification of dynamic flow signals from the static background, but an intermediate region with voxels exhibiting a mixed distribution of dynamic and static scatterers is almost inevitable in practice, which degrades the vascular contrast and connectivity. In this work, the static-dynamic intermediate region was pre-defined according to the asymptotic relation between inverse signal-to-noise ratio (iSNR) and decorrelation, which was theoretically derived for signals with different flow rates based on a multi-variate time series (MVTS) model. Then the ambiguous voxels in the intermediate region were further differentiated using a shape mask with adaptive threshold. Finally, an improved OCTA classifier was built by combining shape, iSNR, and decorrelation features, termed as SID-OCTA, and the performance of the proposed SID-OCTA was validated experimentally through mouse retinal imaging.
What problem does this paper attempt to address?