Polarization-Multiplexed, Dual-Beam Swept Source Optical Coherence Tomography Angiography

Jianlong Yang,Yali Jia,David Huang,Gangjun Liu
DOI: https://doi.org/10.1002/jbio.201700303
2017-01-01
Abstract:A polarization‐multiplexed, dual‐beam setup is proposed to expand the field of view (FOV) for a swept source optical coherence tomography angiography (OCTA) system. This method used a Wollaston prism to split sample path light into 2 orthogonal‐polarized beams. This allowed 2 beams to shine on the cornea at an angle separation of ~14°, which led to a separation of ~4.2 mm on the retina. A 3‐mm glass plate was inserted into one of the beam paths to set a constant path length difference between the 2 polarized beams so the interferogram from the 2 beams are coded at different frequency bands. The resulting OCTA images from the 2 beams were coded with a depth separation of ~2 mm. A total of 5 × 5 mm 2 angiograms from the 2 beams were obtained simultaneously in 4 seconds. The 2 angiograms then were montaged to get a wider FOV of ~5 × 9.2 mm 2 .