PMN-PT Single Crystal Ultrasonic Transducer with Half-Concave Geometric Design for IVUS Imaging

Chunlong Fei,Yaoheng Yang,Feifei Guo,Pengfei Lin,Qiang Chen,Qifa Zhou,Lei Sun
DOI: https://doi.org/10.1109/tbme.2017.2784437
2018-01-01
Abstract:As the key component of intravascular ultrasound (IVUS) imaging systems, traditional commercial side-looking IVUS transducers are flat and unfocused, which limits their lateral resolution. We propose a PMN-PT single crystal IVUS transducer with a half-concave geometry. This unique configuration makes it possible to conduct geometric focusing at a desired depth. To compare performances, the proposed and the traditional flat transducer with similar dimensions were fabricated. We determined that the half-concave transducer has a slightly higher center frequency (35 MHz), significantly broader -6 dB bandwidth (54%) but a higher insertion loss (-22.4 dB) compared to the flat transducer (32 MHz, 28%, and -19.3 dB, respectively). A significant enhancement of the lateral resolution was also confirmed. The experimental results are in agreement with the finite element simulation results. This preliminary investigation suggests that the half-concave geometry design is a promising approach in the development of focused IVUS transducers with broad bandwidth and high lateral resolution.
What problem does this paper attempt to address?