A Fully Integrated 4-Channel GMR Biochip for Biomedical Detection Applications

Cheng Zhu,Lei Zhang,Xizeng Shi,He Qian
DOI: https://doi.org/10.1007/s10470-018-1154-0
IF: 1.321
2018-01-01
Analog Integrated Circuits and Signal Processing
Abstract:This paper proposed a fully integrated 4-channel GMR biochip for biomedical detection assays, including the acquisition analog frontend for small signal extraction, 180° phase shifter, resistor ladder and control circuits for carrier cancellation, and amplifiers. Besides, the overall system was evaluated comprehensively by experiment, and the output noise is as low as \(321.7\mathrm {nV/\sqrt{Hz}},\) which is lower than the signal produced by one single 500nm magnetic particle during the detection.
What problem does this paper attempt to address?