PINK: a tender X‐ray beamline for X‐ray emission spectroscopy

Sergey Peredkov,Nilson Pereira,Daniel Grötzsch,Stefan Hendel,Dirk Wallacher,Serena DeBeer
DOI: https://doi.org/10.1107/s1600577524002200
IF: 2.557
2024-04-27
Journal of Synchrotron Radiation
Abstract:This study showcases the potential of the PINK beamline for advancing research in the field of non‐resonant X‐ray emission spectroscopy in the tender X‐ray energy range from 2.1 keV to 9.5 keV.A high‐flux beamline optimized for non‐resonant X‐ray emission spectroscopy (XES) in the tender X‐ray energy range has been constructed at the BESSY II synchrotron source. The beamline utilizes a cryogenically cooled undulator that provides X‐rays over the energy range 2.1 keV to 9.5 keV. This energy range provides access to XES [and in the future X‐ray absorption spectroscopy (XAS)] studies of transition metals ranging from Ti to Cu (Kα, Kβ lines) and Zr to Ag (Lα, Lβ), as well as light elements including P, S, Cl, K and Ca (Kα, Kβ). The beamline can be operated in two modes. In PINK mode, a multilayer monochromator (E/ΔE ≃ 30–80) provides a high photon flux (1014 photons s−1 at 6 keV and 300 mA ring current), allowing non‐resonant XES measurements of dilute substances. This mode is currently available for general user operation. X‐ray absorption near‐edge structure and resonant XAS techniques will be available after the second stage of the PINK commissioning, when a high monochromatic mode (E/ΔE ≃ 10000–40000) will be facilitated by a double‐crystal monochromator. At present, the beamline incorporates two von Hamos spectrometers, enabling time‐resolved XES experiments with time scales down to 0.1 s and the possibility of two‐color XES experiments. This paper describes the optical scheme of the PINK beamline and the endstation. The design of the two von Hamos dispersive spectrometers and sample environment are discussed here in detail. To illustrate, XES spectra of phosphorus complexes, KCl, TiO2 and Co3O4 measured using the PINK setup are presented.
optics,physics, applied,instruments & instrumentation
What problem does this paper attempt to address?