An Advanced Forward Asymmetric Correction Method for an Electromagnetic Near-Field Probing System
Weiheng Shao,Shan Xue,Xinxin Tian,Liuxing He,Hongyue Wang,Yangyang Zhou,Changjian Zhou,Yiqiang Chen
DOI: https://doi.org/10.1109/tap.2023.3348600
IF: 5.7
2024-02-16
IEEE Transactions on Antennas and Propagation
Abstract:Solving the asymmetric problem is essential for an asymmetric two-port electromagnetic near-field probing system (ATPS). In this article, an advanced forward two-sided asymmetric correction method (FTACM) has been proposed. Unlike the traditional inverse asymmetric calibration method (IACM), it needs two independent electromagnetic excitations created by the calibrator; the solution process of the FTACM is forward and independent of the calibration process. The FTACM only needs the self-parameter of the ATPS, including the -parameter of the symmetric two-port probe and the -parameter of two asymmetric elements. Compared with the IACM, the FTACM has the advantages of no complex mechanical movement (CMM), no symmetry requirement on a calibrator, easy to ultrawideband measurement, greater noise adaptability, and extensible to multiport probing system. A typical calibration result in Case 1 shows that the FTACM can effectively eliminate calibration ripple. The evaluation results of uncertainty prove the reliability of the FTACM, and its maximum uncertainty in Case 2 is 0.12 dB and 1.32° at 18 GHz. The influence and effectiveness of the FTACM on near-field measurement are also analyzed, which shows that the FTACM can correct for asymmetry up to 20 GHz and achieve almost the same characteristics as the original symmetric system.
telecommunications,engineering, electrical & electronic