The Error Analysis and Improvement of Portable Metabolic Rate Detection System

YANG Hao,TANG Fei,LI Shuzhe,WANG Xiaohao,YANG Tao
DOI: https://doi.org/10.3969/j.issn.1004-1699.2014.11.009
2014-01-01
Abstract:In this paper,utilizing Monte Carlo simulation on the error model of portable metabolic rate detection system,obtained the final error indicators. Improved test circuit and used multi-sensor data fusion to achieve metabolic rate test instrument accuracy requirements in non-calibrated conditions. Through sensor error analysis and improvement,the accuracy of temperature test module of the project reached 0. 15 ℃( RMS). In addition,through the maximum likelihood estimate of the improvement circuit for data fusion,it can ensure that the test error is 0. 124 ℃( RMS)or less throughout the temperature testing process,meet the requirements of high accuracy of mass index,avoided the calibration process,reduced the cost of production.
What problem does this paper attempt to address?