Performance Study of a Single Layer THGEM Detector at E3 Line of Beijing Test Beam Facility

Aiwu Zhang,Junguang Lu,Boxiang Yu,Guangpeng An,Jiacai Li,Zhigang Wang,Xiao Cai,Lijun Sun,Jian Fang,Yuguang Xie,Shaoping Zhang,Zunjian Ke,Xilei Sun,Long Gao,Xuan Zhang,Xiaopeng Yang,Chunyang Li,Tao Hu,Li Zhou
DOI: https://doi.org/10.1016/j.nima.2013.04.071
2013-01-01
Abstract:A single layer THick Gaseous Electron Multiplier (THGEM) detector has been studied at E3 line of Beijing Test Beam Facility (BTBF). The responses of the detector to p and pi(+) at momenta 400 MeV/c, 500 MeV/c, 600 MeV/c and 700 MeV/c were investigated at different detector drift distances: 5 mm and 3 mm, the working gas was Neon/CH4 (9515). The p/pi(+) particles were identified through Time of Flight (TOF) method using plastic scintillation detectors. Energy spectra of different particles at various momenta were studied, and they were compared with Geant4 simulation results. Detection efficiencies for p were measured to be about 97% for 5 mm drift distance, and 91% for 3 mm. Detection efficiencies for pi(+) were lower, in the range of 84%-92%. Based on the beam test results, the THGEM detector showed performances satisfying the requirements of a Digital Hadron CALorimeters (DHCAL) in future experiments. (C) 2013 Elsevier B.V. All rights reserved.
What problem does this paper attempt to address?