Operation of a Thick Gas Electron Multiplier (THGEM) in Ar, Xe and Ar-Xe

R Alon,J Miyamoto,M Cortesi,A Breskin,R Chechik,I Carne,J M Maia,J M F dos Santos,M Gai,D McKinsey,V Dangendorf
DOI: https://doi.org/10.1088/1748-0221/3/01/p01005
2008-01-29
Journal of Instrumentation
Abstract:We present the results of our recent studies of a Thick Gaseous Electron Multiplier (THGEM)-based detector, operated in Ar, Xe and Ar:Xe (95:5) at various gas pressures. Avalanche-multiplication properties and energy resolution were investigated with soft x-rays for different detector configurations and parameters. Gains above 104 were reached in a double-THGEM detector, at atmospheric pressure, in all gases, in almost all the tested conditions; in Ar:Xe (95:5) similar gains were reached at pressures up to 2 bar. The energy resolution dependence on the gas, pressure, hole geometry and electric fields was studied in detail, yielding in some configurations values below 20% FWHM with 5.9 keV x-rays.