Operation of MHSP multipliers in high pressure pure noble-gas

F.D. Amaro,J. F. C. A. Veloso,A. Breskin,R. Chechik,J. M. F. dos Santos
DOI: https://doi.org/10.1088/1748-0221/1/04/P04003
2006-01-18
Abstract:We report on the performance of a Micro-Hole & Strip Plate (MHSP) electron multiplier operating in pure Xe, Kr, Ar and Ne at the pressure range of 1 to 6 bar. The maximal gains at 1 bar Xe and Kr are 50000 and 100000, respectively; they drop by about one order of magnitude at 2 bar and by almost another order of magnitude at 5-6 bar; they reach gains of 500 and 4000 at 5 bar in Xe and Kr, respectively. In Ar, the gain varies very little with pressure, being 3000-9000; in Ne the maximum attainable gain, about 100000, is pressure independent above 2 bar. The results are compared with that of single- and triple-GEM multipliers operated in similar conditions. Potential applications are in hard X-ray imaging and in cryogenic radiation detectors.
Instrumentation and Detectors
What problem does this paper attempt to address?
This paper aims to study the performance of the Micro - Hole & Strip Plate (MHSP) electron multiplier in pure inert gases (xenon, krypton, argon and neon) within the pressure range from 1 to 6 bar. Specifically, the paper focuses on the variation of the maximum gain of MHSP under different pressure conditions and compares it with the performance of single - layer and three - layer GEM (Gas Electron Multipliers) multipliers under the same conditions. The main purpose of the research is to evaluate the applicability and performance advantages of MHSP in a high - pressure pure inert gas environment, especially its ability to maintain a relatively high gain at high pressure, which is of great significance for the applications of hard X - ray imaging and cryogenic radiation detectors.