A New Technique for Gaseous Radiation Detectors: The Multigrid High-Pressure Xenon Gas Proportional Scintillation Counter

Filipa I. G. M. Borges,Sergio J. C. do Carmo,João C. R. Mariquito,Alexandre M. F. Trindade,Carlos A. N. Conde
DOI: https://doi.org/10.1109/tns.2010.2052930
IF: 1.703
2010-08-01
IEEE Transactions on Nuclear Science
Abstract:A new technique is presented for high-pressure gaseous radiation detectors leading to pulse amplitudes at least one order of magnitude larger than in ionization-chamber-based gamma-ray detectors. The technique uses room-temperature pure Xe or Xe-based gaseous mixtures at pressures of about 5–20 bar (or higher) to detect ionizing radiation in a multigrid high-pressure gas proportional scintillation counter (MGHP-GPSC) with a CsI deposit as the photocathode in direct contact with the gas, with no optical windows. The detector relies on secondary scintillation as the amplification stage followed by photoelectron production in a CsI photocathode layer. Experimental results are presented for a small prototype filled with pure Xe up to 5 bar, showing that the principle of operation of the detector works. The maximum detector gain obtained so far is about 10.