Measurement of Particle Sulfate from Micro-Aethalometer Filters

Qingqing Wang,Fumo Yang,Lianfang Wei,Guangjie Zheng,Zhongjie Fan,Sanjay Rajagopalan,Robert D. Brook,Fengkui Duan,Kebin He,Yele Sun,Jeffrey R. Brook
DOI: https://doi.org/10.1016/j.atmosenv.2014.07.005
IF: 5
2014-01-01
Atmospheric Environment
Abstract:The micro-aethalometer (AE51) was designed for high time resolution black carbon (BC) measurements and the process collects particles on a filter inside the instrument. Here we examine the potential for saving these filters for subsequent sulfate (SO42-) measurement. For this purpose, a series lab and field blanks were analyzed to characterize blank levels and variability and then collocated 24-h aerosol sampling was conducted in Beijing with the AE51 and a dual-channel filterpack sampler that collects fine particles (PM2.5). AE51 filters and the filters from the filterpacks sampled for 24 h were extracted with ultrapure water and then analyzed by Ion Chromatography (IC) to determine integrated SO42- concentration. Blank corrections were essential and the estimated detection limit for 24 h AE51 sampling of SO42- was estimated to be 1.4 mu g/m(3). The SO42- measured from the AE51 based upon blank corrections using batch-average field blank SO42- values was found to be in reasonable agreement with the filterpack results (R-2 > 0.87, slope = 1.02) indicating that it is possible to determine both BC and SO42- concentrations using the AE51 in Beijing. This result suggests that future comparison of the relative health impacts of BC and SO42- could be possible when the AE51 is used for personal exposure measurement. (C) 2014 Elsevier Ltd. All rights reserved.