Experimental Investigation of Salinity Measurement Based on Optical-Electrical Dual-Differential Methods

Y Zhao,B Zhang,YB Liao,SR Lai
DOI: https://doi.org/10.1117/12.482024
2002-01-01
Abstract:ABSTRACT Experimental setup of a reflex dual-differential sensor for salinity measurement is developed. The sensor exploits thebeam deviation due to optical refraction at the receiving end face of the measurement cell, which is caused by changesin refractive index with different concentration of salt water. With the help of a right-angle reflecting prism and in thepresence of a referenced distilled water cell with an inclined optical window, optical differential measurement ofincident beam deviations is implement, and the deviation caused by changes in salinity will be detected by a PSD withan electrical differential signal processing technique. Experimental setup and measurement principle are discussed.Salinity measurement based on this method avoids the influences by intensity fluctuation of the incident light andtemperature drift. Preliminary experimental results indicate that the estimated salinity measurement resolution can reach O.O2% depending on the performance of PSD. Keywords: Refractive index, salinity measurement, optical inspection, PSD, distilled water, reference
What problem does this paper attempt to address?