Nanometer Al2O3 Microcolumn Preconcentration of Trace Lead and Cadmium in Environment Samples and Their Determination by Graphite Furnace Atomic Absorption Spectrometry

Hong-mei JIANG,Yong-chao QING,Bin HU,Zu-cheng JIANG
DOI: https://doi.org/10.3969/j.issn.1006-6144.2006.02.020
2006-01-01
Abstract:A new method using a microcolumn packed with nanometer Al_(2)O_(3) as solid-phase extractant has been developed for the preconcentration of trace amounts of Pb and Cd prior to their measurements by graphite furnace atomic Absorption spectrometry.Effects of pH,sample flow rate and volume,elution solution and interfering ions on the recovery of the analytes have been investigated.With a concentration factor of 50,the detection limits of the method for Pb and Cd were 0.189 ng/mL and(0.0039 ng/mL,) respectively;and the relative standard deviations were 3.9% and 2.9%,respectively.The method has been applied to the determination of trace Pb and Cd in environmental samples with satisfactory results.