Optimal Design of Optical Length in Low Turbidity Measurement System with Wavelength 1310nm and 1550Nm

Cao Hui-bin,Liu Jian-guo,Gui Hua-qiao,Wang Jie,Wang Huan-qin
DOI: https://doi.org/10.1117/12.2071612
2014-01-01
Abstract:To meet the need of long distance transmission in low turbidity measurement system for low-loss, a new optical structure with wavelength 1310nm and 1550nm as the incident light is employed. In this research, experiments have been done for different optical length of the two wavelength light sources. The results show that: first, the transmitted light intensity has big difference under the circumstance of same concentration and optical length, though the loss has no remarkable difference transmitted in optical fiber between 1310nm and 1550nm. Second, the optimized optical length for better absorbance has been determined for 1310nm and 1550nm and it is irrelevant to the incident intensity. Third, the intensity of the two transmitted light decreases exponentially with the increase of optical length. For example, when the range of the optical length of 1310nm is 0.5mm-2mm, the transmitted intensity is about 60%-79% and the absorbance is 0.12-0.42. The transmitted intensity is about 5%-44%. When the range of the optical length of 1550nm is 0.5mm-2mm and the absorbance is still 0.12-0.42. Our experimental data provides the basis both for the optical length selection of these two light sources in water and the near-infrared spectral wavelength selection.