Reliability of a Compact and Portable Chemiluminescence Detector

Kai-Di Zhang,Xiang-Yu Zeng,Zhi Zeng,Guo-Wei Tao,Jia Zhou
DOI: https://doi.org/10.5875/ausmt.v4i4.845
2014-01-01
International Journal of Automation and Smart Technology
Abstract:Two reliability issues on the compact chemiluminescence detector for glucose measurement based on a single planar transparent EWOD (electrowetting-on-dielectrics) device are studied. Several dielectrics for lowering the manipulation voltage are investigated and 20-27V is realized. An on-chip heater is designed and manufactured for restore the damage of the hydrophobic surface of the EWOD after chemiluminescence reaction. The measurement of glucose shows the sensitivity of the detector reaches 0.12V/mM and the detection range from 1mM to 20mM with a detection limit of 1mM. Such a detector demonstrates its potential as a portable immuno-detector with prompt response and low cost measurement compared with expensive and bulky traditional instruments.