Comparison of intraocular lens power calculation formula

戴锦晖,褚仁远,陆国生
DOI: https://doi.org/10.3969/j.issn.1003-5141.2001.04.023
2001-01-01
Abstract:Objective To compare the accuracy of SRK II formula, SRK/T formula and SCDK formula.Methods A retrospective series of 380 senile cataractous eyes having undergone cataractous surgery were selected for study. The axial length, corneal power, refraction were measured three months later after surgery. The mean absolute refractive error of the three formulas were calculated using the values measured.Results There was no difference in the accuracy of the three formulas when the axial length was less than 26mm.In the group with the axial length above 26mm, the mean absolute refractive error of SRK II formula ,SRK/T formula and SCDK formula were 1.42D,0.84D and 0.89D,respectively. The percentage of the eyes with prediction error greater than 2D was 15.53% in SRK II formula, 4.74% in SRK/T formula and 5.26% in SCDK formula.Conclusion SCDK formula and SRK/T formula are more accurate than SRK II formula when the axial length is above 26mm ,and the accuracy of SCDK formula is the same as SRK/T formula.
What problem does this paper attempt to address?