Presurgery Design and Biomechanical Evaluation of Patient-Specific Knee Implant

Qiguo Rong,Jianfeng Bai,Yongling Huang,Jianhao Lin
DOI: https://doi.org/10.1142/s0219519415400527
2015-01-01
Journal of Mechanics in Medicine and Biology
Abstract:For patients with severe joint deformity or defect, the conventional knee implant cannot meet the installation requirements. A patient-specific prosthesis is a rational choice for these patients. The purpose of this study was to develop a patient-specific knee implant and evaluate its mechanical properties. A mobile-bearing type of knee implant was designed for a young patient with severe rheumatoid arthritis. The stress characteristics were studied by the finite element (FE) method. Results show that the stress distribution of the mobile-bearing knee implant is much better than that of a hinged knee implant designed before. But the stability of the mobile-bearing knee implant should be investigated in more detail before its application.
What problem does this paper attempt to address?