A fiber SPR sensor with high comprehensive evaluation indicator based on core mismatched U-Shaped and tapered arm

Zi-Han Ren,Qi Wang,Xue-Wei Cong,Wan-Ming Zhao,Jing-Ren Tang,Lei Wang,Xin Yan,Ai-Song Zhu,Feng-Mei Qiu,Bin-Hai Chen,Ke-Ke Zhang
DOI: https://doi.org/10.1016/j.measurement.2022.112248
IF: 5.6
2023-01-01
Measurement
Abstract:In the fiber SPR sensor, the resonance superposition caused by many cladding modes and the low intensity ratio of the cladding modes lead to low sensitivity and a broad and shallow resonance spectrum, which is not conducive to high precision sensing. To solve these problems, this paper proposes a core mismatch SPR sensor with tapered and U-shaped structure. The simulation and experimental results show that the introduction of single-mode fiber (SMF) makes the effective sensing area longer, so the resonance valley depth (RVD) increases. And the tapered structure enables cutoff of some higher-order modes, so full width at half maxima (FWHM) becomes narrower. And experiments proved that there is an optimal value of the gold film thickness and macro-bending degree after the introduction of SMF and tapered structure. After parameter optimization, the sensor has 2.5 times higher comprehensive evaluation indicator (CEI, CEI = S*RVD/FWHM) than the U-shaped multimode fiber sensor.
engineering, multidisciplinary,instruments & instrumentation
What problem does this paper attempt to address?