Gas Chromatographic Analysis Of Mixture Of Ethyl Lactate Enantiomer And Ethyl Pyruvate
Tao Guo-Zhong,Yun Guo,Lu Guan-Zhong,Wang Yan-Qin,Guo Yang-Long,Zhi Gang Zhang,Wang Yun-Song,Liu Xiao-Hui
DOI: https://doi.org/10.3321/j.issn:0253-3820.2007.03.031
IF: 1.193
2007-01-01
Chinese Journal of Analytical Chemistry
Abstract:Mixture of chiral enantiomers of ethyl lactate and ethyl pyruvate was separated completely by gas chromatography on chiral stationary phase. The optimized operating parameters were as follows: column temperature 90 degrees C; detector temperature 180 degrees C; injector temperature 200 degrees C; split ratio 1:60; inection volume 0.30 mu L; carrier gas pressure 0. 04 MPa; H-2 pressure 0.10 MPa; air pressure 0.030 MPa, and the retention times of R-,S-ethyl lactate, ethyl pyruvate were 13.0 min, 14.02 min, 7.96 min, respectively. The contents of R, S-ethyl lactate enantiomer and ethyl pyruvate were quantitatively determined by using cyclopentanone as an internal standard with relative correction factor 2.145, 2.139, 2.805, respectively. The results show that this method with relative error of -1.7% - + 1.2%, recovery of 97.03% - 101.9% and relative standard deviation of 0.2% - 0.9% is simple, highly efficient, accurate, stable, well repeatable and widely linear for determining the content of ethyl pyruvate, R-and S-ethyl lactate in the mixture solution by GC and internal standard method.