Determination Of Metalaxyl Enantiomeric Purity By Nonchiral High Performance Liquid Chromatography With Circular Dichroism Detector

Chen Shengwen,Lin Kunde,Xi Meng,Liu Weiping
DOI: https://doi.org/10.3321/j.issn:0253-3820.2006.04.021
IF: 1.193
2006-01-01
Chinese Journal of Analytical Chemistry
Abstract:A novel method for the determination of enantiomeric purity of chiral compounds by nonchiral high performance liquid chromatography (HPLC) coupled with circular dichroism, (CD) detector was developed. The principle of the method was proposed and illustrated by the analysis of enantiomeric purity of metalaxyl. The results indicated that g values (anisotropy factor) had a good linearity with enantiomeric excess (ee) and correlation coefficient was 0.9974. The relative average deviations between the ee evaluated by the developed method and those determined by chiral HPLC were less than 3.0%. The proposed method has a high accuracy and shows a good perspective to determine the ee of chiral compounds.
What problem does this paper attempt to address?