Anodic Stripping Voltammetric Measurement of Trace Cadmium at Tin-Coated Carbon Paste Electrode

Bang Lin Li,Zhi Ling Wu,Chang Hong Xiong,Hong Qun Luo,Nian Bing Li
DOI: https://doi.org/10.1016/j.talanta.2011.11.070
IF: 6.1
2011-01-01
Talanta
Abstract:A carbon paste electrode modified with tin was used for the determination of trace cadmium by anodic stripping voltammetry. The electroanalytical performance for the determination of Cd(II) on the tin-coated carbon paste electrode(SnF-CPE) was better than that on the carbon paste electrode. The measuring conditions have been optimized. The measurement of trace cadmium on the SnF-CPE has the best response under the conditions of 0.10molL−1 acetate buffer solution (pH 3.9), 3.5mgL−1 Sn(II), deposition potential of −1.40V, and deposition time of 150s. The SnF-CPE revealed highly linear behavior in the concentration range of 2.0–90.0μgL−1 with the detection limit of 1.13μgL−1 for Cd(II). The developed sensor has been applied to the determination of Cd(II) in real water samples with satisfactory results.
What problem does this paper attempt to address?