High Sensitivity and Reproducibility of a Bismuth/Poly(bromocresol Purple) Film Modified Glassy Carbon Electrode for Determination of Trace Amount of Cadmium by Differential Pulse Anodic Stripping Voltammetry

Gongjun Yang,Jiankang Yan,Fen Qi,Cheng Sun
DOI: https://doi.org/10.1002/elan.201000260
IF: 3.077
2010-01-01
Electroanalysis
Abstract:This work described a novel type of bismuth/poly(bromocresol purple) film modified glassy carbon electrode (denoted as Bi/Poly(BCP)/GCE) for anodic stripping analysis of trace Cd2+. The Bi/Poly(BCP)/GCE was fabricated in situ by depositing simultaneously bismuth and cadmium by reduction at -1.20 V on the poly(BCP) film using a differential pulse voltammetry. Under the optimum conditions, the anodic stripping peak current response increased linearly with the Cd2+ concentrations in a range of 2.0 x 10(-8)-1.0 x1.0(-7) M and 1.0 x 10(-7)-6.0x10(-6) M in 0.1 M NaAc-Ac buffer solution (pH 5.0) with the detection limit of 6.5 x 10(-9) M (S/N =3). The Bi/poly(BCP)/GCE performed good reproducibility and high sensitivity. Finally, this proposed method was successfully applied to determine the concentration of Cd2+ in water samples.
What problem does this paper attempt to address?