Variable slew-rate spiral design: theory and application to peak B(1) amplitude reduction in 2D RF pulse design.

Dan Xu,Kevin F. King,Zhi-Pei Liang
DOI: https://doi.org/10.1002/mrm.21375
2007-01-01
Magnetic Resonance in Medicine
Abstract:A new class of spiral trajectories called variable slew-rate spirals is proposed. The governing differential equations for a variable slew-rate spiral are derived, and both numeric and analytic solutions to the equations are given. The primary application of variable slew-rate spirals is peak B, amplitude reduction in 2D RF pulse design. The reduction of peak B, amplitude is achieved by changing the gradient slew-rate profile, and gradient amplitude and slew-rate constraints are inherently satisfied by the design of variable slew-rate spiral gradient waveforms. A design example of 2D RF pulses is given, which shows that under the same hardware constraints the RIF pulse using a properly chosen variable slew-rate spiral trajectory can be much shorter than that using a conventional constant slew-rate spiral trajectory, thus having greater immunity to resonance frequency offsets.
What problem does this paper attempt to address?