In situ impedance measurements in diamond anvil cell under high pressure.

Yue Wang,Yonghao Han,Chunxiao Gao,Yanzhang Ma,Cailong Liu,Gang Peng,Baojia Wu,Bao Liu,Tingjing Hu,Xiaoyan Cui,Wanbin Ren,Yan Li,Ningning Su,Hongwu Liu,Guangtian Zou
DOI: https://doi.org/10.1063/1.3282444
IF: 1.6
2010-01-01
Review of Scientific Instruments
Abstract:Two-electrode configuration was developed for in situ electrical impedance detecting on diamond anvil cell under high pressure. The metal gasket was used as one electrode and the risk coming from electrical short between sample and interside wall of the gasket was eliminated. The configuration was evaluated and proved to be effective by measuring the electric impedance of nanocrystalline ZnS under high pressure.