Online spectrophotometric determination of Fe(II) and Fe(III) by flow injection combined with low pressure ion chromatography.

Shujuan Chen,Nan Li,Xinshen Zhang,Dongjing Yang,Heimei Jiang
DOI: https://doi.org/10.1016/j.saa.2014.11.071
2014-01-01
Abstract:A simple and new low pressure ion chromatography combined with flow injection spectrophotometric procedure for determining Fe(II) and Fe(III) was established. It is based on the selective adsorption of low pressure ion chromatography column to Fe(II) and Fe(III), the online reduction reaction of Fe(III) and the reaction of Fe(II) in sodium acetate with phenanthroline, resulting in an intense orange complex with a suitable absorption at 515nm. Various chemical (such as the concentration of colour reagent, eluant and reductive agent) and instrumental parameters (reaction coil length, reductive coil length and wavelength) were studied and were optimized. Under the optimum conditions calibration graph of Fe(II)/Fe(III) was linear in the Fe(II)/Fe(III) range of 0.040-1.0mg/L. The detection limit of Fe(III) and Fe(II) was respectively 3.09 and 1.55μg/L, the relative standard deviation (n=10) of Fe(II) and Fe(III) 1.89% and 1.90% for 0.5mg/L of Fe(II) and Fe(III) respectively. About 2.5 samples in 1h can be analyzed. The interfering effects of various chemical species were studied. The method was successfully applied in the determination of water samples.
What problem does this paper attempt to address?