Diagnosis of Labral Tears with MR Arthrography Three-Dimensional Isotropic Volumetric Interpolated Breath-Hold Examination Sequence

TIAN Chun-yan,ZHENG Zhuo-zhao
DOI: https://doi.org/10.13929/j.1003-3289.2011.12.042
2011-01-01
Abstract:Objective To observe the diagnostic value of shoulder MR arthrographic fat-suppressed three-dimensional(3D) isotropic volumetric interpolated breath-hold examination(VIBE) sequence for labral tears.Methods MR arthrographic fat-suppressed 3D isotropic VIBE and conventional T1W turbo spin-echo sequences were performed in 79 patients.The results of arthroscopy were regarded as gold standards,the sensitivity,specificity and accuracy of fat-suppressed 3D VIBE and conventional TSE images for detection of labral tears were calculated and compared by 2 radiologists,respectively.K values were calculated to quantify the level of interobserver agreement.Results Twenty-seven of the 79 patients were classified as SLAP lesions,44 as anterior labral tears,and 10 as posterior labral tears.There was no statistical difference between fat-suppressed 3D isotropic VIBE and conventional TSE regarding the sensitivity(81.48%—88.89%,85.19%—92.59%) and specificity(82.69%—84.62%,84.62%—86.54%) of SLAP lesions,sensitivity(88.64%—90.91%,90.91%—95.45%) and specificity(94.29%—97.14%,97.14%) of anterior labral tears,nor sensitivity(80.00%—90.00%,80%) and specificity(92.75%—95.65%,94.20%—95.65%) of posterior labral tears.At interobserver comparison,the consistency was very good.Conclusion Fat-suppressed 3D isotropic VIBE MR arthrography is a reliable method to evaluate labral tears.