MR Arthrographic 3D VIBE Sequence for Bankart Lesions at 3.0-T

Zhuozhao Zheng
2012-01-01
Abstract:Objective To evaluate the diagnostic value of shoulder MR arthrographic three dimensional(3D) volumetric interpolated breath hold examination(VIBE) sequence in Bankart lesions.Materials and Methods Water excitation 3D VIBE and conventional T1-weighted turbo spin-echo sequences were performed in 99 patients with shoulder MR arthrography.According to the gold standard——the results of arthroscopy,sensitivity and specificity of 3D VIBE and conventional T1-weighted images for detection of Bankart lesions and bony Bankart lesions were calculated and compared using paired McNemar test.K values were calculated to quantify the level of interobserver agreement.Results 58 of the 99 patients were classified as Bankart lesions,including 26 bony Bankart lesions at arthroscopy.Other 41 were found with intact anterior-inferior labrum.The sensitivity and specificity had no statistically significant differences of Bankart lesions between 3D VIBE and conventional T1-weighted images(91.38%-93.10% versus 91.38%-96.55% and 95.12%-97.56% versus 97.56%).3D VIBE was significantly higher sensitivity than conventional T1-weighted images in detecting bony Bankart lesions(96.15% versus 42.31%-50%),but not with specificity(95.89%-97.26% versus 97.26%-100%).The interobserver agreements were excellent.Conclusion MR arthrographic 3D VIBE is a comprehensive method for evaluating Bankart lesions.