Biofilm Thickness and Temperature Sensor Based on Fiber Bragg Grating

LUO Bin-bin,ZHAO Ming-fu,ZHOU Xiao-jun,HUANG De-yi,WANG Shao-fei,CAO Xue-mei
DOI: https://doi.org/10.16136/j.joel.2011.06.001
2011-01-01
Abstract:A biofilm sensor based on the single-end etched fiber Bragg grating(FBG) is proposed and is used for simultaneous measurement of biofilm thickness and temperature in the biofilm reactor.The theoretical model for the simultaneous measurement of surrounding refractive index(SRI) and temperature by the single-end etched FBG is presented,and the linear approximation for the SRI sensitivity is given.A single-end etched FBG with the diameter of the etched region of approximately 7.1-7.4 μm was fabricated.Experimental results demonstrate that the sensor shows a good linear property and SRI sensitivity of 6.894±0.029 nm/riu in the refractive index range from 1.333 to about 1.355,approximately agreeing with the simulated value of 6.83 nm/riu.Also,by using a spectral unit with wavelength resolution of 1 pm,the resolutions of 15.36±0.85 μm and 0.12 ℃ for the biofilm thickness and temperature are realized respectively.
What problem does this paper attempt to address?