Application of SPR Detector for Flow Analysis System

CHEN Hui-bin,WANG Xiao-ping,WU Hong-wei,ZHU Yan,LI Ya-ping
DOI: https://doi.org/10.3785/j.issn.1008-973x.2011.08.026
2011-01-01
Abstract:To develop a novel detector with characteristics of versatility,low sample consumption,and inexpensiveness,surface plasma resonance(SPR) detector was designed for flow analysis system.Its performance was evaluated using 3 glucide solutions..Results showed that the RSD values were below 1.3%,and the detection limits of D-xylose,sucrose and glucose were 92,123,and 132 ng,respectively.The linearities of three components were excellent between detection limits and 10000 mg/L with correlation coefficients above 0.999.The SPR detector was applied to the liquid chromatography system.The RSD values were below 2.3% for separated samples with standard concentration solutions,it could detect the sucrose and galactose as low as 6 and 9 μg,respectively.It indicates that the detector is preponderant to detect materials insensitive to the ultraviolet detection or fluorescence detection,or which only can be detected by derivatization.