Abstract:Traditional liquid chromatography detectors don't have both strong currency and high sensitivity. According to this,an SPR liquid chromatography detector including SPR sensor,micro flow cell,circuit and software was developed.A chromatography system was put up using this detector together with pump,separation column and sampler to detect carbohydrate.A chromatography of a mixed solution of xylose,glucose and sucrose was obtained.The detection limit of glucose was 20μg,linear range was 20~160μg.The results showed that SPR detector has strong currency,but its sensitivity needs to be further improved.