Experimental Investigation and Numerical Simulation of the Hydraulic Performance of a Centrifugal Blood Pump

王鑫,朱雷,罗先武,刘树红,曹广军
DOI: https://doi.org/10.16511/j.cnki.qhdxxb.2011.04.023
2011-01-01
Abstract:External characteristic tests were made with a radial blood pump by adjusting the motor speed and controlling the flow on a close-loop experimental rig to analyze the relationship between the flow in the blood pump and the hydro performance and biocompatibility.Three-dimensional steady turbulent flow was then simulated to compare with the test results.The results indicate that the head of the blood pump is more than 1 m with a flow discharge of 5 L·min^-1,with the blood pump operating in a wide range.The results show that the Reynolds number effect cannot be neglected at low rotational speeds.Flow characteristic analysis shows that the wall shear stress in the region near the trailing edge of the suction side on the vane surface is relatively large,thus forming dangerous regions for blood cells.
What problem does this paper attempt to address?