Design for reliability of ultrasonic bone knife

Ying Chen,Rui Kang,Zhaoying Zhou
DOI: https://doi.org/10.1166/asl.2011.1685
2011-01-01
Advanced Science Letters
Abstract:Design for reliability (DfR) of ultrasonic bone knife starts at the early design phase and can greatly improve the safety of surgery which is especially crucial for patient's life. By failure analysis of ultrasonic bone knife, the main failure mechanism is found. DfR measures are taken to improve the design, which include improving shape discontinuity of knife transition step, adopting arc teeth instead of saw teeth and changing location of dangerous cross-section. We also present the Finite Element Analysis (FEA) based virtual qualification process for the DfR of ultrasonic bone knife. Resonant frequency and knife tip vibration amplitude that calculated by ANSYS are measured and verified by dynamic signal analyzer and digital microscope. Stress distribution of dangerous parts before and after DfR is compared. Improvement of stress concentration was achieved by DfR measures, which is in accordance with lifetime statistical data.
What problem does this paper attempt to address?