Matching Technology Analysis between the Front-end Electronics and Neutron Detectors Array System

Lian CHEN,Feng LI,Xiao JIANG,Fu-tian LIANG,Ge JIN
DOI: https://doi.org/10.3969/j.issn.0258-0934.2010.05.029
2010-01-01
Abstract:In the ICF experiment, it should keep a certain distance between the neutron detectors and electronics equipments for eliminating neutron radiation, it should be considered carefully for a long distance transmission for the signal of detector because of attenuation of the long coaxial cable. In addition, the signal of X-rays is very large, more than 50 V, the signal of secondary neutron which we are interested in is normally below 1 V, and the time interval between two signals is about 240 ns. The reflection, saturation, overshoot induced by the large X-ray may affect the measurement of the following neutron signal. Focused on the problems of signal transmission and receiving in long distance, the studies have been done with experimental test. It is showed that the characteristic of signal can be kept with coaxial cable and the X-ray signal can be suppress without affecting the measurement of neutron based the circuits we are used.