Development of a Submillimeter Double-Ridged Waveguide Ortho-Mode Transducer (OMT) for the 385–500 GHz Band

Mamoru Kamikura,Masato Naruse,Shin’ichiro Asayama,Naohisa Satou,Wenlei Shan,Yutaro Sekimoto
DOI: https://doi.org/10.1007/s10762-010-9632-1
2010-01-01
Abstract:We present design and evaluations of a submillimeter double-ridged waveguide ortho-mode transducer (OMT) for ALMA Band 8 (385–500 GHz) cartridge receiver. The measured transmission loss of the OMT at 4 K was 0.4–0.5 dB according to noise measurements with an SIS mixer. The polarization isolation was measured to be larger than 29 dB from quasioptical measurements. The OMT consists of a Bϕifot junction and a double-ridged guide. A robust design with allowable mechanical errors of 20 μm has been demonstrated.
What problem does this paper attempt to address?