EVALUATION OF SMALL ANIMAL IMAGING UNDER DIFFERENT SCAN CONDITIONS

贾鹏翔,王浩宇,闫镔,李磊,陈健,张锋,包尚联
DOI: https://doi.org/10.3969/j.issn.1005-1120.2010.01.005
2010-01-01
Abstract:s To evaluae small animal imaging with individual different high voltage,filter thickness and tube cur-rent,an animal X-ray micro-computed tomography (micro-CT) system based on panel detector is developed and a rat is scanned by using the system with individual high voltage,tube current,filter thickness,and exposure time.A model is presented based on the Monte Carlo code PENELOPE for generating the X-ray spectra of X-ray tube used in the micro-CT system.A platform developed based on Matlab allows for calculating beam quality parame-ters,including the average energy of X-ray beam,the change of transmition rate and the input X-ray fluence.The factors affecting the signal difference to noise ratio (SDNR) of micro-CT are investigated and the relationship between SDNR and scan combinations is analyzed.A series of tools and methods are developed for small animal imaging and imaging performance evaluation in the field of small animal imaging.